AID series alarm and display indicator are remote display and audible and visible alarm devices. AID duplicates fault, alarm and operating message of monitoring devices. When fault occurs, AID gives audible and visible signal to medical staff and technical staff. Additionally, response value of AIM-M series monitoring device can be set by AID remotely. AID Series alarm and display device have three types: AID10, AID120 and AID150, which shall be arranged in operating room or nursing station so that it can be permanently monitored by medical staff.
Aux Power |
Voltage |
DC 24V |
Power Consumption |
≤0.6W |
|
Insulation Measure Range |
10~999kΩ |
|
Insulation Alarm Range |
50~999kΩ |
|
Transformer Load Rate |
Percentage display |
|
Current Alarm Range |
14A、18A、22A、28A、35A、45A |
|
Temperature Alarm Range |
0~+200℃ |
|
Alarm Mode |
Audible and visual alarm |
|
Alarm Type |
Insulation fault, overload, over temperature, device failure |
|
Communication Mode |
RS485(MODBUS-RTU) |
|
Display Mode |
LED display |